Field Metrology Well
Fluke Calibration 9172
Display accuracy
Dry-wells
are typically calibrated by inserting a calibrated PRT into one of the
wells and making adjustments to the calibrator’s internal control sensor
based on the readings from the PRT. This has limited value because the
unique characteristics of the reference PRT, which essentially become
“calibrated into” the calibrator, are often quite different from the
thermometers tested by the calibrator. This is complicated by the
presence of significant thermal gradients in the block and inadequate
sensor immersion into blocks that are simply too short.
Metrology
Wells are different. Temperature gradients, loading effects, and
hysteresis have been minimized to make the calibration of the display
much more meaningful. We use only traceable, accredited PRTs to
calibrate Metrology Wells and our proprietary electronics consistently
demonstrate repeatable accuracy more than ten times better than our
specs, which range from ±0.1 °C at the most commonly used temperatures
to ±0.25 °C at 661 °C.
For
even better accuracy, Metrology Wells may be ordered with built-in
electronics for reading external PRTs with ITS-90 characterizations.
(See sidebar, Built in Reference Thermometry)